Home Subscription Services
 
   

 
Quintessence International
QI Home Page
About the Editor
Editorial Board
Accepted Manuscripts
Submit
Author Guidelines
Submission Form
Reprints / Articles
Permissions
Advertising
MEDLINE Search
 
 
 
 
 
FacebookTwitter
Quintessence Publishing: Journals: QI
Quintessence International

Edited by Eli Eliav

ISSN 0033-6572 (print) • ISSN 1936-7163 (online)

Publication:
June 1993
Volume 24 , Issue 6

Back
Share Abstract:

Precision of hand-held dental radiometers

Rueggeberg

Pages: 391-396
PMID: 8234644

The intensity of light within the wavelengths of 400 to 500 nm on a dental photocurable resin restorative material is a vital factor affecting many of the physical and chemical properties of the resulting restoration. This paper compares the precision of two recently introduced hand-held radiometers. The results indicate that the response of both hand-held meters was linear. For a given amount of decrease in light source intensity, both hand-held units responded with a corresponding decrement in measured intensity value. Both hand-held meters were found to limit their pass band of intensity readings between 400 and 515 nm, making them very useful clinical instruments. However, the absolute intensity readings between the two units were significantly different.

Full Text PDF File | Order Article

 

Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.
  © 2014 Quintessence Publishing Co Inc
 

Home | Subscription Services | Books | Journals | Multimedia | Events | Blog
Terms of Use | Privacy Policy | About Us | Contact Us | Advertising | Help | Sitemap | Catalog