Precision of hand-held dental radiometers


Pages: 391-396
PMID: 8234644

The intensity of light within the wavelengths of 400 to 500 nm on a dental photocurable resin restorative material is a vital factor affecting many of the physical and chemical properties of the resulting restoration. This paper compares the precision of two recently introduced hand-held radiometers. The results indicate that the response of both hand-held meters was linear. For a given amount of decrease in light source intensity, both hand-held units responded with a corresponding decrement in measured intensity value. Both hand-held meters were found to limit their pass band of intensity readings between 400 and 515 nm, making them very useful clinical instruments. However, the absolute intensity readings between the two units were significantly different.

Full Text PDF File | Order Article


Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.
  © 2015 Quintessence Publishing Co Inc

Home | Subscription Services | Books | Journals | Multimedia | Events | Blog