LOGIN
 
Share Page:
Back

Volume 14 , Issue 2
March/April 1999

Pages 233238


Bone-Hydroxyapatite Interface in Retrieved Hydroxyapatite-Coated Titanium Implants: A Clinical and Histologic Report

Adriano Piattelli, MD, DDS, Antonio Scarano, DDS, Luca Di Alberti, DDS, Maurizio Piattelli, MD, DDS


PMID: 10212540

Two hydroxyapatite-coated implants were retrieved after 12 months of loading because of a fracture of the abutments. The specimens were treated to obtain thin ground sections, and a microprobe chemical analysis was done under a scanning electron microscope equipped with an energy-dispersive x-ray analysis and cathodoluminescence system. Under light microscopy, close contact between the bone and the hydroxyapatite coating was seen, with no gaps at the interface. In some areas of the coating a reduction of the coating thickness could be observed, along with the presence of some detached hydroxyapatite particles embedded in newly formed bone. The chemical analysis of the cathodoluminescent areas at the interface showed a reduced calcium:phosphorus ratio in this region. (Int J Oral Maxillofac Implants 1999;14:233238) Key words: cathodoluminescence, hydroxyapatite, interface, osseointegration, titanium implants


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2014 Quintessence Publishing Co, Inc JOMI Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Accepted Manuscripts
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help