Volume 25 , Issue 3
May/June 2010

Pages 532–539

Parameters of Implant Stability Measurements Based on Resonance Frequency and Damping Capacity: A Comparative Finite Element Analysis

Werner Winter, Dr-Ing/Stefan Möhrle, Dipl-Ing/Stefan Holst, Priv-Doz, Dr Med Dent/Matthias Karl, Priv-Doz, Dr Med Dent

PMID: 20556252

Purpose: Contradictory results have been reported on the comparability of implant stability measurements performed with the Periotest and the Osstell Mentor devices. The purpose of the present finite element analysis was to simulate the influence of the parameters implant length, bone quality (cortical thickness and damping factor), bone loss, and quality of transducer fixation on resonance frequency analysis (RFA) and damping capacity measurements. Materials and Methods: Three-dimensional finite element models of implants placed in human mandibular bone were designed for the simulation of Periotest (Periotest value; PTV) and RFA (implant stability quotient) measurements. Three values for each of the parameters implant length, damping capacity of cortical and trabecular bone, thickness of cortical bone, bone loss, and quality of transducer fixation were obtained. Measurements were simulated at four stages of osseointegration. Results: For all parameters, an increase in implant stability was found with increasing levels of osseointegration. Implant stability was positively correlated with implant length and thickness of cortical bone, with slightly converging values at increased levels of osseointegration. Varying the damping factor of bone had no significant effect. Implant stability was negatively correlated with bone loss, with slightly converging values at increased levels of osseointegration. Linear changes in implant length and bone loss caused nonlinear effects in implant stability values. Stiffness of transducer fixation had an impact on RFA measurements when values below 10 GPa were applied. Conclusion: Although both measuring devices reacted similarly when different parameters of implant stability were changed, good correlation between Periotest values and implant stability quotients was observed only when measurement values of implants without bone loss were considered. Int J Oral Maxillofac Implants 2010;25:532–539

Full Text PDF File | Order Article


Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.


© 2015 Quintessence Publishing Co, Inc JOMI Home
Current Issue
Ahead of Print
Author Guidelines
Accepted Manuscripts
Submission Form
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us