Accuracy of Mechanical Torque- Limiting Devices for Implants

Jon P. Standlee, DDS, Angelo A. Caputo, PhD, Ming-ya J. Chwu, DDS, Tao T. Sun, DDS


PMID: 11958404

Purpose: To examine the accuracy of 3 mechanical torque wrenches. Materials and Methods: The torque outputs of the Nobel Biocare, Straumann ITI, and DynaTorq ITL mechanical torque-limiting devices were determined using a special setup on an Instron test machine. The devices were held in the test setup and oriented so that activation of the drivers caused a pure torsion effect. Results: Significant differences generally existed between individual units and the target torque levels for the Nobel Biocare torque controller. Discussion: The mean torque values of the ITI and ITL devices were within 10% of their respective target torque levels. Knowledge of applied torque levels to the screws that retain implant abutments and their attached prostheses is necessary to achieve optimal preload. The ITI and ITL devices tested in this study were capable of providing consistent torque at or near their respective targets. Conclusion: The torque output of each individual device deviated in varying degrees from target torque values. (INT J ORAL MAXILLOFAC IMPLANTS 2002;17:220224)


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2014 Quintessence Publishing Co, Inc JOMI Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Accepted Manuscripts
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help