The Effect of Luting Agents on the Retention and Marginal Adaptation of the CeraOne Implant System
George H. Clayton, DDS/Carl F. Driscoll, DMD/Stephen O. Hondrum, DDS, MS
PMID: 9337028
In this study, various luting agents were evaluated to determine their retentive strengths as they pertain to the CeraOne single-tooth implant system. Ten samples of five different luting agents (zinc oxide–eugenol, glass-ionomer cement, hybrid glass-ionomer cement, composite resin, and zinc phosphate) were tested for retentive strength of the CeraOne gold cylinder to the CeraOne abutment. Under the conditions of the experiment, zinc phosphate showed a mean retentive strength 164% greater than that of glass-ionomer cement and 49% greater than that of composite resin cement. Scanning electron micrographs were taken to evaluate the effect of various luting agents on marginal opening. The measurements revealed that zinc phosphate had the greatest marginal opening, although its mean value of 62 µm is within clinically acceptable limits. (INT J ORAL MAXILLOFAC IMPLANTS 1997;12:660–665)
Key words: CeraOne, luting agents, marginal opening, retention