Torque Generated by Handheld Screwdrivers and Mechanical Torquing Devices for Osseointegrated Implants
Kevin L. Goheen, DDS/Stanley G. Vermilyea, DMD, MS/Jafar Vossoughi, PhD/ John R. Agar, DDS, MA
PMID: 8206549
This study evaluated the ability of practitioners experienced in the use of Brånemark implant components to impart a desired torque using handheld screwdrivers. The torque output and variability of mechanical torque devices was also examined. For clinicians using handheld screwdrivers, values ranged from 0.7 to 18.1 Ncm, 1.4 to 33.7 Ncm, and 8.2 to 36.2 Ncm for the 10-Ncm, 20-Ncm, and 32-Ncm torques respectively. The manually operated mechanical torquing devices produced torque values within the tolerances specified by their respective manufacturers at all levels. Significant variation from the manufacturers set values were noted as a function of tightening speed with the electronic torquing device. The data indicate that there is wide variation in the ability of clinicians to perceive adequate torquing forces applied to implant components. Calibrated torquing devices are mandatory if proper torquing procedures are to be accomplished. (INT J ORAL MAXILLOFAC IMPLANTS 1994;9:149—155)
Key words: abutment screws, gold screws, preload, screw loosening, torque meter, yield strength