LOGIN
 
Share Page:
Back

Volume 2 , Issue 1
Spring 1987

Pages 11-14


Force Transfer by Osseointegration Implant Devices

Mark E. Kinni, D.M.D./Stephanie N. Hokama/ Angelo A. Caputo, Ph.D.


PMID: 3552975

The load distributing characteristics of two implant designs, the Brånemark and Core-Vent systems, were studied under both vertical and inclined loading using photoelastic models. Applied loads were transferred to the supporting structure more equitably with the Brånemark design than with the Core-Vent design.