Resistance of Internal-Connection Implant Connectors Under Rotational Fatigue Loading

H.W. Anselm Wiskott, DMD, MS, MSD, PD / Robin Jaquet, DMD / Susanne S. Scherrer, DMD, PD / Urs C. Belser, DMD, Prof Dr Med Dent


PMID: 17465350

Purpose: To aid in developing mechanically optimized implant-abutment connectors, the fatigue resistance of 5 connector configurations of the Replace Select system (Easy abutment, Easy abutment without antirotational mechanism, Multi-unit abutment, Esthetic Alumina abutment, Esthetic Zirconia abutment) was investigated. Other purposes of the study were to determine whether the connector’s antirotational mechanism participates in fatigue resistance and to compare the results with previous data on Straumann connectors. Materials and Methods: The repetitive, alternating, and multivectorial intraoral force pattern was reproduced by subjecting the test specimens to the rotating cantilever beam test. To this end, the samples were spun around their long axis while clamped into a revolving collet on one end and loaded normal to their long axis on the other end. The aim was to determine the load level at which 50% of the specimens survived and 50% fractured before 106 cycles. Means were determined using the staircase procedure. They were fitted with 95% confidence intervals for intergroup comparisons. Results: In the chosen testing configuration, 2 statistical groups emerged. The Easy abutments with and without antirotational mechanism were statistically similar, with mean failure loads in the 70 to 72 N range. Both ceramic and the Multi-unit abutments belonged to the second group, with mean failure loads in the 53 to 58 N range. Conclusions: (1) The fatigue resistance of ceramic and the Multi-unit abutments was approximately 20% less than that of the Easy Abutments. (2) The antirotational mechanism did not participate in mechanical resistance. (3) The fatigue strength of the Easy abutment connectors was approximately 20% greater than the equivalent abutments in the Straumann system. Int J Oral Maxillofac Implants 2007;22:249–257

Key words: connectors, dental implants, failure, fatigue


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2014 Quintessence Publishing Co, Inc JOMI Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Accepted Manuscripts
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help