In Vitro Evaluation of Three Methods for the Retrieval of Fractured Screw Fragments from Dental Implants


Rubén Agustín-Panadero, DDS, PhD/Carlos Labaig-Rueda, MD, DDS, PhD/Beatriz Castillo-Rodriguez, DDS/Alberto Ferreiroa, DDS, PhD/María Fernanda Solá-Ruíz, MD, DDS, PhD


PMID: 28231345
DOI: 10.11607/jomi.5174

Purpose: To compare the effectiveness of different methods of fractured screw fragment retrieval from dental implants, comparing a conventional method (use of an exploration probe and ultrasonic tips) with two mechanical retrieval kits. Materials and Methods: One hundred thirty-five screws in implant internal connections were fractured to test fragment retrieval comparing three methods: conventional method with the use of an explorer probe and ultrasonic tips (group CE), extraction with a mechanical kit with a guide cylinder (group MK1), and extraction with a mechanical kit without a guide cylinder (group MK2). Extraction success and retrieval times for each method, as well as success and retrieval times in relation to fracture depth—coronal, middle, or apical—were analyzed by means of chi-square, Fisher exact, Kruskal-Wallis, and Mann-Whitney tests (P < .05). The integrity of implant internal threads was also analyzed. Results: The extraction success rates for groups MK2, CE, and MK1 were 93.3%, 73.3%, and 20%, respectively. The mean ± SD extraction times were 2 minutes, 56 seconds ± 2 minutes, 9 seconds; 3 minutes, 15 seconds ± 1 minute, 48 seconds; and 13 minutes, 34 seconds ± 7 minutes, 25 seconds for groups MK2, CE, and MK1, respectively. Statistically significant differences were found in the number of fractured screws extracted and in the retrieval time. No significant differences were found for retrieval time or extraction success in relation to the depth of the fracture. Conclusion: The mechanical kit without the guide cylinder achieved the best results with respect to the number of screw fragments retrieved, retrieval time, and preservation of implant internal threads.


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2017 Quintessence Publishing Co, Inc JOMI Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Accepted Manuscripts
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help