Surface Analysis and Effects on Interfacial Bone Microhardness of Collagen-Coated Titanium Implants: A Rabbit Model

Marco Morra, Dr Chem/Clara Cassinelli, Dr Biol/Laura Meda, Dr Phys/Milena Fini, MD/Gianluca Giavaresi, MD/Roberto Giardino, MD


PMID: 15747670

Purpose: The aim of this study was to evaluate the surface chemistry and the microhardness at the implant-bone interface using a recently developed collagen-coated titanium implant in a short-term rabbit model. Materials and Methods: Surface chemistry was evaluated by x-ray photoelectron spectroscopy (XPS), while in vivo studies involved 4-week implants mid-diaphysis in the lateral femurs of adult male rabbits. After conventional embedding and evaluation of histologic sections, the resin-embedded blocks containing the implanted screws were used to measure bone hardness by means of an indentation test. Results: Decomposition of the C1s peak obtained by XPS analysis confirmed that surface-immobilized collagen retained all the molecular features of the control, nonimmobilized reference. As to microhardness measurement, newly formed bone at the collagen-coated-implant/bone interface was significantly harder than bone at the interface of the uncoated control implant and bone. Discussion: These results suggested that collagen coating significantly improves bone maturation and mineralization at the interface in comparison with uncoated commercially pure titanium. Surface modification of titanium implants by collagen coating has recently been discussed as a promising approach to the biochemical modification of implant surfaces. The present results support previous histologic findings and demonstrated that the biomolecular layer linked over the titanium implant can increase the bone healing rate, at least in this animal model. Conclusions: The present microhardness measurement at the bone-implant interface showed that collagen coating can significantly improve bone maturation and mineralization at the interface in comparison with uncoated commercially pure titanium, confirming and substantiating previous findings by histomorphometric measurements from the same model. Int J Oral Maxillofac Implants 2005;20:2330


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2014 Quintessence Publishing Co, Inc JOMI Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Accepted Manuscripts
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help