The Effect of a Combination of Implant Controller and Handpiece from Different Manufacturers on the Torque Value

Du-Hyeong Lee, DDS, MS/Yong-Gun Kim, DDS, PhD/Jong-Ho Lee, DDS, PhD/Sam-Pyo Hong, DDS, PhD/Young-Jun Lim, DDS, PhD/Kyu-Bok Lee, DDS, PhD


PMID: 26252027
DOI: 10.11607/jomi.3553

Purpose: To determine the accuracy of applied torque of different implant controller and handpiece combinations by using an electronic torque gauge. Materials and Methods: Four combinations of the following devices were tested: Surgic XT controller (NSK), XIP10 controller (Saeshin), X-SG20L handpiece (NSK), CRB26LX handpiece (Saeshin). For five torque settings, 30 measurements were recorded at 30 revolutions per minute by using an electronic torque gauge fixed to jigs, and means were calculated. Results: Applied torques were generally higher than the set torque of 10 and 20 Ncm and lower than the set values of 40 and 50 Ncm. The average torque deviations differed significantly among the combinations (P < .05). At 10 and 20 Ncm, the Surgic XT/X-SG20L combination yielded the closest value to the intended torque, followed by the XIP10/X-SG20L combination. At 30 Ncm, the XIP10/X-SG20L combination showed the nearest value. At 40 Ncm, the Surgic XT/X-SG20L, XIP10/CRB26LX, and XIP10/X-SG20L combinations showed deviations within 10%. At 50 Ncm, all the combinations showed lower applied torque than the set value. Large standard deviations were observed in the Surgic XT/CRB26LX (13.288) and Surgic XT/X-SG20L (7.858) combinations. Conclusion: Different combinations of implant controllers and handpieces do not generate significant variations in applied torque. The actual torque varies according to the torque setting. It is necessary to calibrate devices before use to reduce potentially problematic torque.


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2017 Quintessence Publishing Co, Inc JOMI Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Accepted Manuscripts
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help