Inter- and Intraobserver Variability in Resonance Frequency Analysis of Palatal Implants—A Technical Note

Michael H. Bertl, DMD/Rüdiger Emshoff, MD, DMD/Aleš Celar, MD, DMD/Adriano G. Crismani, DMD

PMID: 24066338
DOI: 10.11607/jomi.2831

Purpose: To assess inter- and intraobserver variability in wireless resonance frequency analysis (RFA) stability measurements of palatal implants and to evaluate the influence of age, sex, time after implant insertion, and measurement direction on variability. Materials and Methods: Three observers conducted wireless RFA stability measurements of palatal implants from 16 patients. Measurements were taken in anteroposterior and laterolateral directions and were repeated after 1 hour. Results: Data showed a small interobserver variation (1.93) with intraobserver variation (1.77) as its largest component. Time after implantation showed a strong influence (P = .027) on the interobserver variation. Conclusions: The tested wireless RFA device can be considered a reliable instrument for measuring the stability of palatal implants. Multiple readings and observers could further improve measurement reliability.