Volume 27 , Issue 4
July/August 2012

Pages e49-e57


Biomechanical Effect of a Zirconia Dental Implant–Crown System: A Three-Dimensional Finite Element Analysis


Chih-Ling Chang, DDS, MS/Chen-Sheng Chen, PhD/ Tze Cheung Yeung, BDS, MS/Ming-Lun Hsu, DDS, Dr Med Dent


PMID: 22848900

Purpose: The objective of this study was to analyze and compare the stresses in two different bone-implant interface conditions in anisotropic three-dimensional finite element models (FEMs) of an osseointegrated implant of either commercially pure titanium or yttrium-partially stabilized zirconia (Y-PSZ) in combination with different superstructures (gold alloy or Y-PSZ crown) in the posterior maxilla. Materials and Methods: Three-dimensional FEMs were created of a first molar section of the maxilla into which was embedded an implant, connected to an abutment and superstructure, using commercial software. Two versions of the FEM were constructed; these allowed varying assignment of properties (either a bonded and or a contact interface), so that all experimental variables could be investigated in eight groups. Compact and cancellous bone were modeled as fully orthotropic and transversely isotropic, respectively. Oblique (200-N vertical and 40-N horizontal) occlusal loading was applied at the central and distal fossae of the crown. Results: Maximum von Mises and compressive stresses in the compact bone in the two interfaces were lower in the zirconia implant groups than in the titanium implant groups. A similar pattern of stress distribution in cancellous bone was observed, not only on the palatal side of the platform but also in the apical area of both types of implants. Conclusion: The biomechanical parameters of the new zirconia implant generated a performance similar to that of the titanium implant in terms of displacement, stresses on the implant, and the bone-implant interface; therefore, it may be a viable alternative, especially for esthetic regions. Int J Oral Maxillofac Implants 2012;27:e49–e57

Key words: dental implants, finite element method, stress, zirconia


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2013 Quintessence Publishing Co, Inc JOMI Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Accepted Manuscripts
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help