Stress Distribution on Scalloped Implants with Different Microthread and Connection Configurations Using Three-Dimensional Finite Element Analysis


Kyung-soo Choi, DDS, MS, PhD/Seong-hun Park, PhD/Jae-hoon Lee, LBD/Young-chan Jeon, DDS, MSD, PhD/Mi-jung Yun, DDS, MSD/Chang-mo Jeong, DDS, MSD, PhD


PMID: 22616069

Purpose: The objective of this study was to investigate the effects of different microthread designs and implant-abutment connection configurations of scalloped implants on stress distribution in bone using three-dimensional finite element analysis. Materials and Methods: Three different designs of scalloped implants (two different microthread designs and one without microthreads) with two different connection systems and two flat-top implants with similar connection systems were compared in a bone model that mimicked the anterior maxilla. Vertical and oblique (30-degree) loading with 100 N of force was applied to eight models. Peak stress levels and the distribution of stress were observed. Results: The stress pattern of scalloped implants was distinctively different from that observed on flat-top implants. Scalloped implants showed peak stresses in the proximal cortical bone as well as in the buccal and palatal cortical bone, whereas flat-top implants showed peak stresses mainly in the buccal and palatal cortical bone and limited stress in the proximal bone. The scalloped implant without microthreads and a conical platform-switched connection demonstrated the lowest peak stress levels. The scalloped implant with a straight platform connection generally showed peak stress that was two to three times higher than that seen in the conical platform-switched model. Conclusions: Peak stress levels in scalloped implants varied with microthread designs, connection configurations, and the direction of loading. The conical platform-switched connection seemed more important for a scalloped implant than the microthread design in reducing loading stresses exerted on the surrounding bone. Scalloped implants without microthread and a with a conical platform-switched connection or closed microthreads and a conical platform-switched connection showed consistently lower buccal bone stress than the flat-top implants in areas where the bone had a sloping and scalloping shape. INT J ORAL MAXILLOFAC IMPLANTS 2012;27:e29e38.


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2014 Quintessence Publishing Co, Inc JOMI Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Accepted Manuscripts
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help