LOGIN
 
Share Page:
Back

Volume 9 , Issue 5
September/October 1996

Pages 452-458


Scanning Electron Microscopy and Profilometer Evaluation of Glazed and Polished Dental Porcelain

Fuzzi/Zaccheroni/Vallania


PMID: 9108746

The surface roughness of Vita VMK porcelain following oven glazing and eight grinding/polishing treatments was analyzed qualitatively using scanning electron microscopy and quantitatively using a profilometer. Techniques were selected to simulate clinical practice. Scanning electron microscopy evaluation found oven glazing produced a better surface than other polishing methods. On the basis of the profilometric examination, the best roughness average value was obtained using diamond instruments with progressively smaller particle sizes (30, 15, and 8 um). Scanning electron microscopy analysis showed that all the treatments left the surfaces partially prous and cracked; however, the glazed surface yielded the best result. Although no significant differences were detected for the different treatments, the use of a 30-um diamond instrument pr oduced a rougher surface. No correlation was found between scanning electron microscopy and profilometer results. This study suggests that several procedures may be used to effectively finish ceramic surface


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2017 Quintessence Publishing Co, Inc

IJP Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Accepted Manuscripts
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help