LOGIN
 
Share Page:
Back

Volume 9 , Issue 5
September/October 1996

Pages 419-433


The Spline Implant: Design, Engineering, and Evaluation

Binon


PMID: 9108742

The design and engineering characteristics of the Spline dental implant were investigated. Compressive strength were evaluated. The results of the mechanical test indicated that this implant/abutment complex is mechanically stable, has minimal rotational movement, improved resistance to screw loosening, and good interface fidelity. The data for the new design are compared with published data for other implant/abutment connections.