LOGIN
 
Share Page:
Back

Volume 12 , Issue 3
May/June 1999

Pages 249-254


Investigations into the failure of dental magnets.

Riley MA, Williams AJ, Speight JD, Walmsley AD, Harris IR


PMID: 10635193

PURPOSE: The aim of this study was to determine the mechanisms of failure of magnets used for denture retention. MATERIALS AND METHODS: Magnets were retrieved from dentures that had failed after 4 to 18 months of clinical use. The magnets were sectioned and the failure modes were evaluated using optical or scanning electron microscopy. RESULTS: The magnetic retention system investigated consists of a neodymium-iron-boron magnet sandwiched in a stainless steel keeper, capped with a titanium dome, and sealed with epoxy resin. The diffusion of moisture through the epoxy seal and the breakdown of the encapsulating material led to corrosion of the magnet. In both cases, a loss in retention occurred as the corrosion products weakened the overall magnetic attraction. CONCLUSION: The failure of magnets may occur by either breakdown of the epoxy seal or breakdown of the encapsulating material. Both problems need to be addressed to prolong the life span of these magnets.


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2017 Quintessence Publishing Co, Inc

IJP Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Accepted Manuscripts
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help